• Always hang the nurse control pendant onto a foot end siderail or the linen tray (option) before removing the footboard.
• Do not raise the lower leg section while the bed extender is in use. This is to avoid the situation where the product does
not support the lower legs of a taller patient.
• Do not use the siderails to move the product. Always move the product using the integrated handles in the headboard
and footboard.
• Always place the patient control pendant safely on the mattress while the pendant is in use.
• Always hang the patient control pendant onto the siderail when the pendant is not in use.
• Do not squeeze or pinch the pendant cord in the bed frame.
• Always place the nurse control pendant onto the footboard.
• Always hang the nurse control pendant onto a foot end siderail or store in the linen tray (option) before removing the
footboard.
• Do not squeeze or pinch the pendant cable in the bed frame.
• Do not remove the footboard after you extend the bed extender.
• The safe working load of the linen tray is 15 kg.
• Only use authorized accessories for this product. The use of unauthorized accessories may result in product damage or
injury to the operator or patient. Stryker is not responsible for any damage or injury that may result from the misuse of
the product or the use of unauthorized accessories.
• Always make sure that accessories are locked into position.
• Do not allow accessories to interfere with mechanical or electric mechanisms of the product.
• Always remove the lifting pole before transporting the product.
• Do not use the lifting pole as a push or pull device.
• Do not use the oxygen bottle holder as a push or pull device.
• Always turn the oxygen bottle holder in toward the bed before transporting a patient.
• Do not strike the oxygen bottle holder while transporting a patient.
• Do not load the oxygen bottle holder above the safe working load of 7.5 kg.
• The safe working load of each Foley hook is 2 kg.
• Do not steam clean, pressure wash, ultrasonically clean, or immerse any part of the product in water. Exposure to water
may damage the internal electric parts. These methods of cleaning are not recommended and may void this product's
warranty.
• Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning. Some
cleaning products are corrosive in nature and may cause damage to the product if you use them improperly. If you do
not properly rinse and dry the product, you may leave a corrosive residue on the surface of the product that could cause
premature corrosion of critical components. Failure to follow these cleaning instructions may void your warranty.