Télécharger Imprimer la page

Stryker SV2 Manuel D'utilisation page 47

Masquer les pouces Voir aussi pour SV2:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 419
E E M M C C i i n n f f o o r r m m a a t t i i o o n n
G G u u i i d d a a n n c c e e a a n n d d M M a a n n u u f f a a c c t t u u r r e e r r ' ' s s D D e e c c l l a a r r a a t t i i o o n n - - E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c E E m m i i s s s s i i o o n n s s
S S V V 2 2 is intended for use in an electromagnetic environment specified below. The customer or the user of S S V V 2 2 should
E E m m i i s s s s i i o o n n s s T T e e s s t t
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
N N o o t t e e : : The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR
11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
S S V V 2 2 is suitable for use in the electromagnetic environment specified below. The customer or the user of S S V V 2 2 should
I I m m m m u u n n i i t t y y T T e e s s t t
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrostatic fast Transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
KK-7500 REV 08
assure that it is used in such an environment.
G G u u i i d d a a n n c c e e a a n n d d M M a a n n u u f f a a c c t t u u r r e e r r ' ' s s D D e e c c l l a a r r a a t t i i o o n n - - E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c I I m m m m u u n n i i t t y y
assure that it is used in such an environment.
I I E E C C 6 6 0 0 6 6 0 0 1 1 T T e e s s t t L L e e v v e e l l
+8 kV contact
+15 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV line(s) to line(s)
+2 kV line(s) to earth
C C o o m m p p l l i i a a n n c c e e
Group 1
Class A
Class A
Complies
C C o o m m p p l l i i a a n n c c e e L L e e v v e e l l
+8 kV contact
+15 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV line(s) to line(s)
+2 kV line(s) to earth
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c E E n n v v i i r r o o n n m m e e n n t t
S S V V 2 2 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
S S V V 2 2 is suitable for use in all
establishments other than domestic
and those directly connected to the
public low voltage power supply
network that supplies buildings used
for domestic purposes.
E E l l e e c c t t r r o o m m a a g g n n e e t t i i c c
E E n n v v i i r r o o n n m m e e n n t t G G u u i i d d a a n n c c e e
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality is that
of a typical commercial and/
or hospital environment.
43
EN

Publicité

loading

Ce manuel est également adapté pour:

7500