Data entry of premeasured
sensors
Calibration / adjustment
Data entry of premeasured sensors
Data entry of premeasured sensors
Entry of cell constant and zero point of a sensor, related to 25°C,
1013 mbars.
During calibration the analyzer is in function check mode.
Current outputs and relay contacts behave as configured.
Menu Display
Calibration
Module 8285 El. COND
Module 8285 El. COND
Module 8285 pH/ORP
Return
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
Data entry
Temperature
Conductivity
Sensor replacement
Cell constant
End
Data entry
premeasured sensors
0.225 mS/cm
25.0°C
Info
0.225 mS/cm
25.0 °C
0.225 mS/cm
25.6 °C
+25.6°C
0.225 mS/cm
1.000
Select module: COND
During calibration, the output
currents (1 and 2), limit contacts,
and controller output are in function
check mode.
Confirm with enter.
Select calibration method
"Data entry"
Confirm with enter.
Enter cell constant
of premeasured sensor
Confirm with "OK"
or repeat calibration.
35