Calibrating the sensors
Calibration
Calibrating the sensors
Sensor calibration
Since the cell constant is subject to production-related variances, the
dismounted sensor should be calibrated with a calibration solution
(e.g. NaCl saturated).
The cell constant of the sensor – particularly of a fringe-field sensor –
depends on the type of installation:
•
When the sensor is mounted in a free space (minimum distances
exceeded), the cell constant can be entered directly as given in the
specifications. Calibration method: "Data entry"
•
When mounted in restricted space (minimum distances not kept), the
sensor must be calibrated when mounted since the resulting cell constant
has changed. Calibration method: "Product calibration"