Menu Display
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
Manual entry
Dip sensor in cal solution
then `Start` calibration
Measured cal temp
Sensor replacement
Return
Manual entry
Calibration running
Determine pair of cond/°C values
Calibration temperature
Response time
Manual entry
Enter calibration solution
for correct temperature!
Measured cal temp
Conductivity
Repeat
Calibration data record
Calibration
Cal mode
Cal temperature
Conductivity
Cell constant
End
32
Manual entry of cal solution
0.020 mS/cm
25.0 °C
0.020 mS/cm
24.9 °C
+24.9°C
Start
0.020 mS/cm
24.9 °C
+24.9°C
0001s
0.020 mS/cm
25.0 °C
+025.0 °C
0251 mS/cm
End
0250 mS/cm
25.6 °C
21.01.04 11:37
Manual input
+025.6°C
0249 mS/cm
2.7450 /cm
Adjust
Select calibration menu
Select "Module COND"
Select calibration method
"Manual entry of cal solution".
Confirm with enter.
Enter process temperature, if manual
temperature adjustment has been
selected.
Immerse sensor in cal solution.
Start calibration with softkey or
enter.
Calibration is running.
The display shows:
Calibration temp
Response time
Enter conductivity.
Press "End" softkey to
terminate calibration.
Adjustment
Press "Adjust" to take over the
values determined during calibration
for calculating the measured
variables.