Menu Display
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
Automatic
Dip sensor in cal solution
then "Start" calibration.
Calibration solution
TC automatically considered
Enter cal temp
Sensor replacement
Return
Automatic
Calibration running
Correction of cell constant
Calibration temp
Solution table value
Response time
End
Calibration data record
Calibration
Cal mode
Cal temperature
Conductivity
Cell constant
End
30
Automatic calibration
0250 mS/cm
Select calibration menu
25.0 °C
Select "Module COND"
Select calibration method "Automatic
with standard cal solution".
Confirm with enter.
0250 mS/cm
25.0 °C
NaCl saturated
+025.0 °C
Start
0250 mS/cm
25.6 °C
+025.0 °C
0.020 mS/cm
0003s
Repeat
Adjustment
0250 mS/cm
25.6 °C
Press "Adjust" to take over the
values determined during calibration
21.01.04 11:37
Automatic
for calculating the measured
+025.6°C
0249 mS/cm
variables.
2.7450 /cm
Adjust
Display of selected calibration
solution.
Enter process temperature, if manual
temperature adjustment has been
selected.
Dip sensor in calibration solution.
Start calibration with softkey or
enter.
Calibration is running.
The display shows:
•
Calibration temp
•
Solution table value
(conductivity in dependence on
cal temperature)
•
Response time