Menu Display
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
Product calibration
Cal medium: product
Cal by taking sample
and entering conductivity
without temp compensation
Return
Product calibration
Step 1: Sampling
‚Save' the sample value
‚Input' lab value
Conductivity
Temperature
Input
Product calibration
Step 2: Lab value
Input sample lab value
Lab value
Sensor replacement
Abort
Calibration data record
Calibration
Cal mode
Cal temperature
Conductivity
Cell constant
End
34
0249 mS/cm
25.0 °C
0249 mS/cm
25.0 °C
Start
0249 mS/cm
25.0 °C
0.000µS/cm
25.0 °C
Save
0249 mS/cm
25.0 °C
0.249mS/cm
OK
0250 mS/cm
25.6 °C
21.01.04 11:37
Product calibration
+025.6°C
0249 mS/cm
2.7450 /cm
Adjust
Product calibration
Select calibration menu
Select "Module COND"
Select calibration method
"Product calibration".
Confirm with enter.
Step 1
Take sample.
Store measured value and tempe-
rature at the moment of sampling
("Save" softkey or enter).
The analyzer will automatically
return to Calibration mode selection.
Press meas to return to measure-
ment.
Exception:
Sample value can be measured on
the site and be entered immediately.
To do so, press "Input" softkey.
Step 2
Lab value has been measured. When
you open the Product calibration
menu again, the display shown on
the left appears: Enter reference
value ("Lab value"). Confirm with
"OK" or repeat calibration.
Adjustment
Press "Adjust" to take over the
values determined during calibration
for calculating the measured
variables.
8285