To calibrate a COND module: Select a calibration method
(1) Press the menu key to access menu selection
(2) Pressing the meas key returns to measurement
(3) Select Calibration menu group using the arrow keys
(4) Press enter to confirm, enter passcode
(5) Select COND module, confirm with enter
(6) Select calibration method
Calibration
Maintenance Parameter setting Diagnostics
1147
4
Module I
Module II
Module III
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
Menu groups
3
5
6
0.020 mS/cm
25.0 °C