Relay contacts

Relay contacts
Parameter setting/Module BASE/Relay contacts
Menu Display
Contact K1 (Administrator)
Usage
Variable
Limit value
Hysteresis
Effective direction
Contact type
Return
BASE module
Alarm (K4)
K1
K2
K3
Contact assignment: See terminal plate of BASE module
64
7.00 pH
19.2°C
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output
The BASE module provides 4 relay contacts
(max. AC/DC rating 30 V / 3 A each).
Contact K4 is provided for failure message.
The switching behavior (normally open or
normally closed), as well as a switch-on or
switch-off delay can be defined.
Default settings of the user-definable relay
contacts of the BASE module:
K3:
K2:
K1:
The contact assignment K1 - K3 is user defined
("Usage"):
NAMUR maintenance request
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output (COND module only)
Setting the relay contacts

Relay contacts, usage

Stromausgang parametrieren
• Parametrierung aufrufen
Call up parameter setting
• Paßzahl eingeben
Enter passcode
• Modul BASE wählen
Select "Module BASE"
• "Augangsstrom I1" auswählen
Select "Contact ..."
"Usage" (Fig.)
NAMUR maintenance request
NAMUR function check
Limit value