Setting the message parameters
Messages
Note: Function check active
Menu Display
Messages (Administrator)
Messages Conductivity
Messages Resistivity
Messages Concentration
Messages Temperature
Messages Salinity
Return
Messages pH value (Administrator)
Monitoring
Abort
Messages (Administrator)
Monitoring
Failure Limit Lo
Warning Limit Lo
Warning Limit Hi
Failure Limit Hi
Abort
Message list
Fail Temperature range
Fail pH range
Fail Meas. processing
Abort
56
0.220 mS/cm
20.1 °C
0.220 mS/cm
20.1 °C
Off
Device limits max.
Variable limits
OK
0.220 mS/cm
20.1 °C
Variable limits
–02.00 pH
–02.00 pH
+16.00 pH
+16.00 pH
OK
0.220 mS/cm
22.3 °C
Messages
Messages
All parameters determined by the
measuring module can generate
messages.
• Device limits max:
Messages are generated when the
process variable (e.g. conductivity)
is outside the measurement range.
The "Failure" icon is displayed, the
NAMUR failure contact is activated
(BASE module, factory setting:
contact K4, N/C contact).
The current outputs can signal a
22 mA message (user defined).
• Variable limits:
For the "failure" and "warning"
messages you can define upper and
lower limits for message generation.
• Message icons:
Failure (Failure limit HiHi/LoLo)
Maintenance (Warning limit Hi/Lo)
Diagnostics menu
When the "Maintenance" or
"Failure" icons are flashing in the
display, you should call up the
Diagnostics menu. The messages are
displayed in the "Message list".
8285