Switching parameter sets via OK2
Parameter setting / System control / Function control matrix
Note: Function check active
Parameter sets
2 complete parameter sets (A, B) can be stored in the analyzer.
You can switch between the parameter sets using the OK2 input.
The currently activated set can be signaled by a relay contact. An icon in the
measurement display shows which parameter set is active:
or
Menu Display
Function control matrix (Administrator)
Input OK2
Left softkey
Right softkey
Profibus DO 2
Return
Contact K3 (Administrator)
Usage
Contact type
ON delay
OFF delay
Abort
Note
The selection has no effect when working on SmartMedia card with
SW 8285-102.
7.00pH
24.8 °C
ParSet KI rec.
Fav Unical
–
–
–
–
–
–
Connect
7.00pH
24.8 °C
NAMUR maintenance
NAMUR function check
Limit
Rinse contact
Parameter set B active
USP output
OK
Parameter sets
Select parameter set (A, B) via
input OK2
•
Call up parameter setting
•
System control
•
Function control matrix
•
Select "OK2"
•
Connect "Parameter set A/B"
Signal active parameter set
via relay contact
•
Call up parameter setting
•
Module BASE
•
Select contact
•
Usage: "Parameter set ...".
67