Parameter setting: Lock functions

Administrator level: Enable / lock functions for Operator level
Note: Function check active (Parameter setting: Module BASE)
Menu Display
Parameter setting (Administrator)
System control
Module 8285 FRONT
Module 8285 BASE
Module 8285 pH/ORP
Module 8285 pH/ORP
Module 8285 El. COND
Return
Module 8285 pH/ORP (Administrator)
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return
Module 8285 pH/ORP (Administrator)
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return
Module 8285 pH/ORP
Input filter
Sensor data
Cal preset values
TC process medium
ORP/rH value
Delta function
Return
38
Administrator level:
Enable / lock functions
Example:
Blocking access to the calibration
adjustments from the Operator level
11.03 pH

Call up parameter setting

25.0°C
Select Administrator level.
Enter passcode (1989).
Select "Module pH/ORP" (e.g.)
using arrow keys,
confirm with enter.
Select "Cal preset values"
11.03 pH
25.0°C
using arrow keys.
"Block" with softkey.
Blocked
Now, the "Cal preset values" line is
11.03 pH
25.0°C
marked with the "lock" icon. This
function cannot be accessed from the
Operator level any more. The softkey
function changes to " Enable".
Release
Call up parameter setting
11.03 pH
25.0°C
Select Operator level,
enter passcode (1246).
Select "Module pH/ORP" (e.g.).
Now, the locked function is displayed
in gray and marked with the "lock"
icon.
8285