Parameter setting
TC process medium
Note: Function check active
Menu Display
Module 8285 El. COND (Administrator)
Input filter
Sensor data
Cal preset values
TC process medium
Messages
Return
TC process medium (Administrator)
TC
Abort
TC process medium (Administrator)
EN 27888: Natural waters (0 ... 35 °C)
Ultrapure water: traces of impurity
TC correction
Reference temperature
Impurity
Abort
0.245
Outp I1 5.70 mA
0.220 mS/cm
20.1 °C
Blocked
0.220 mS/cm
20.1 °C
Off
Linear
EN 27888
Ultrapure water
OK
0.220 mS/cm
20.1 °C
NaOH
NaCl
HCl
NH3
OK
TC
mS/cm
20.1
°C
Favorites menu
TC process medium
TC process medium
You can choose from:
• Linear (input of TC coefficient)
• EN 27888
• Ultrapure water
(Additional function SW 8285-008)
When you have selected "Ultrapure
water", you must specify the type of
impurity:
NaOH
Alkaline ultrapure water
NaCl
Neutral ultrapure water, for
conductivity measurement in water
processing behind gravel bed filter
HCl
Acidic ultrapure water, for conduc-
tivity measurement behind cation
filter
NH 3
Ammoniacal ultrapure water
When the TC correction for pro-
cess medium is switched on, "TC"
appears in the display in measuring
mode.