Boston Scientific LithoVue Manuel D'utilisation page 40

Masquer les pouces Voir aussi pour LithoVue:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 96
User Message
Problem with LithoVue™
Flexscope icon is displayed in
video field.
Hardware Malfunction icon is
displayed in the video field.
LithoVue Flexscope Use time
Exceeded icon is displayed in
the video field.
the System Workstation does
not start when you press the
power button.
40
Black (K) ∆E ≤5.0
Possible cause
LithoVue Flexscope
malfunction, missing or
corrupt calibration data.
LithoVue Flexscope
malfunction.
System Workstation internal
hardware communication
failure.
LithoVue Flexscope use
time exceeded.
Power Transformer's power
switch in the off position.
The Power Transformer's
AC power cord not properly
connected to power
transformer, or the AC Power
Cord is not connected tightly to
the hospital's electrical outlet.
The Power Transformer's DC
Power Cord is not connected
tightly to the back of the
Touch PC.
No electrical power at the AC
power receptacle.
Defective power transformer.
corrective Action
Remove the LithoVue Flexscope. Ensure
the connection terminals are dry and
clean. Ensure the plug on the Flexscope
Cable is properly aligned with the key on
the Workstation Receptacle. Ensure the
plug is fully inserted into the receptacle.
If User Message remains after a repeated
connection, replace the LithoVue Flexscope.
Remove the LithoVue Flexscope. Ensure
the connection terminals are dry and
clean. Ensure the plug on the Flexscope
Cable is properly aligned with the key on
the Workstation Receptacle. Ensure the
plug is fully inserted into the receptacle.
If User Message remains after a repeated
connection, proceed to step below.
Reboot the System Workstation. If the
problem still exists, contact Boston
Scientific for support.
Dispose of the LithoVue Flexscope and
replace with a new one.
Ensure the Power Transformer's switch is
turned on and the green LED is lit.
Ensure the ends of the AC Power Cord are
firmly connected to the Power Transformer
and to a power source.
Ensure the end of the DC Power Cord is
firmly connected to receptacle on the
Touch PC.
Check the hospital's electrical outlet.
Contact Boston Scientific using the
information found in Section 8 "Service
and Warranty".
User's Manual

Publicité

Table des Matières
loading

Table des Matières