Installation and electrical wiring
Fig. 12 :
Installation of a wall-mounted version of the 8025
4.2.

Electrical connection

4.2.1.
Recommendations for the electrical wiring (all
versions)
Do not open nor wire while the device is powered.
The electrical installation of the building where the device is installed must be
secured by a switch or a circuit breaker. Install it near the device and identify it as
the switchgear of the device.
Do not apply in the same cable both a dangerous voltage and a safety extra low
voltage to the relays.
Use safety devices for both:
-
The power supply: a time-delay fuse (300 mA) and an interrupter
-
Relays: a max. 3-A-fuse and a circuit breaker (depending on the application).
Use cables with an operating limit temperature of at least 80 °C.
Earth the device locally by using a wire with a cross section of at least 0,75 mm
Under normal operating conditions the measuring signal can be transmitted by a shielded
cable with 0,75 mm
The line must not be installed in combination with carrying lines with a higher voltage
or frequency; If a combined installation cannot be avoided, a minimum space of 30 cm
Fuse
2
cross-section.
E-18-
Flow transmitter 8025/8035
115/230 VAC voltage supply