Stryker Neurovascular TransForm Mode D'emploi page 7

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 14
Table 7. average TransForm™ Occlusion Balloon Catheter Deflation Time
(for Super Compliant Sizes) (Seconds)
Viscosity (cp)
Balloon Size
Brand Name
@ 37˚C
Omnipaque
240
3.4
®
3x5
Omnipaque 300
6.3
Omnipaque 350
10.4
Omnipaque 240
3.4
4x7
Omnipaque 300
6.3
Omnipaque 350
10.4
Omnipaque 240
3.4
4x10
Omnipaque 300
6.3
Omnipaque 350
10.4
Omnipaque 240
3.4
7x7
Omnipaque 300
6.3
Omnipaque 350
10.4
Omnipaque 240
3.4
7x10
Omnipaque 300
6.3
Omnipaque 350
10.4
Omnipaque 240
3.4
7x15
Omnipaque 300
6.3
Omnipaque 350
10.4
warranTy
Stryker Neurovascular warrants that reasonable care has been used in the design and
manufacture of this instrument. This warranty is in lieu of and excludes all other warranties
not expressly set forth herein, whether express or implied by operation of law or otherwise,
including, but not limited to, any implied warranties of merchantability or fitness for a
particular purpose. Handling, storage, cleaning and sterilization of this instrument as well
as other factors relating to the patient, diagnosis, treatment, surgical procedures and other
matters beyond Stryker Neurovascular's control directly affect the instrument and the results
obtained from its use. Stryker Neurovascular's obligation under this warranty is limited to the
repair or replacement of this instrument and Stryker Neurovascular shall not be liable for
any incidental or consequential loss, damage or expense directly or indirectly arising from
the use of this instrument. Stryker Neurovascular neither assumes, nor authorizes any other
person to assume for it, any other or additional liability or responsibility in connection with
this instrument. Stryker neurovascular assumes no liability with respect to instruments
reused, reprocessed or resterilized and makes no warranties, express or implied, including
but not limited to merchantability or fitness for a particular purpose, with respect to such
instruments.
Transend™ EX Platinum, Transend EX Floppy, Transend EX Soft Tip, Transend EX Standard, Transend 300 ES,
Transend 300 Floppy guidewires are manufactured for Stryker by Boston Scientific.
Ethiodol and Lipiodol are trademarks of Guerbet Group.
Omnipaque is a trademark of GE Healthcare.
Contrast/Saline Dilution
50/50
67/33
100% Contrast
5
5
6
5
6
10
5
6
14
7
9
14
8
10
19
9
11
25
7
9
14
8
10
23
10
14
43
20
28
47
25
34
85
28
42
148
24
32
58
28
40
102
37
51
204
30
42
71
36
55
134
44
62
249
7