This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that it is used in such an environment.
Emissions
Compliance
according to
RF Emissions
Group 2
(CISPR 11)
CISPR B
Class B
Emissions
Classification
The System is intended for use in the electromagnetic environment specified
below. You should observe these requirements when using the System.
Immunity
IEC 60601-1-2
against
test level
ElectroStatic
contact
Discharge, ESD
discharge:
(IEC 61000-4-2)
± 8 kV
air discharge:
± 15 kV
Power
30 A/m
frequency
magnetic fields
50/60 Hz (IEC
61000-4-8)
Radiated RF
10 V/m at
(IEC 61000-4-3)
80 MHz–
2.7 GHz
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflections from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (mobile/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the equipment is used
exceeds the applicable RF compliance level above, the equipment should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orientating or relocating the equipment.
184

Electromagnetic Emissions

Electromagnetic environment
The Pod and the PDM emit low-level
electromagnetic energy (RF) in order to
communicate. Although unlikely, nearby
electronic equipment may be affected.
The System is suitable for use in all establishments
including domestic establishments.

Electromagnetic Immunity

Compliance
level (of this
device)
Electromagnetic environment
If floors are covered with synthetic material,
try to avoid electrostatic discharges.
± 8 kV
± 15 kV
400 A/m
Suitable for most environments. Magnetic
field strengths in excess of 400 A/m would
be unlikely except in close proximity to
industrial magnetic devices.
10 V/m
Suitable for most environments. Keep
portable RF communications equipment
at least 30 cm (12 inches) away from the
Omnipod DASH™ System.