10 Alarms, Notifications, Communication Errors
Error when deactivating a Pod
If a communication error occurs during Pod deactivation, the PDM offers these
options:
•
DISCARD POD: Select this option to tell your PDM to unpair from that Pod.
The PDM instructs you to remove your Pod and tap CONTINUE.
•
TRY AGAIN: Select this option to attempt to re-establish communication.
Note: After selecting the discard option, you can prevent future alarms from the
discarded Pod by following the instructions in "Silencing an Alarm" on page 123.
Note: If there is an unconfirmed bolus when you discard a Pod, the PDM does
not know how much of the bolus was delivered. Therefore, the PDM temporarily
disables the Bolus Calculator. If you tap the Bolus button while the Bolus
Calculator is disabled, the PDM displays a message that says 'Bolus Calculator
temporarily disabled'. You can deliver a manually calculated bolus when the Bolus
Calculator is disabled.