Pod Specifications - OmniPod DASH Guide De L'utilisateur

Masquer les pouces Voir aussi pour DASH:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Duration of Insulin
Action
Maximum bolus size
Extended Bolus
Bolus preset
Suspend
Low reservoir volume
advisory
Pod expiry notification 1 to 24 hours in 1-hour increments. Default is 4 hours.
Auto-off timer
History screen display
Language

Pod Specifications

Size: 3.9 cm wide x 5.2 cm long x 1.45 cm high (1.53" x 2.05" x 0.57")
Weight (without insulin): 26 grams (0.92 oz)
Operating temperature range: Pod operating environment of 5 °C to 40 °C
(41 °F to 104 °F).
Note: The Pod temperature equilibrates from 23 °C to 37 °C (73 °F to 98.6
°F) when worn on the body.
Start-up temperature: above 10 °C (50 °F)
Storage temperature range: 0 °C to 30 °C (32 °F to 86 °F)
Reservoir volume (deliverable): 200 units
Cannula insertion depth: 4-7 mm (0.16-0.28 in)
Depth of insulin infusion: ≥ 4 mm (0.16 in)
Waterproof rating: IP28 (7.6 metres (25 feet) for up to 60 minutes)
Insulin concentration: U-100
Alarm type: Audible. Output: ≥ 45 db(A) at 1 metre
Operating relative humidity range: 20 to 85%, non-condensing
Storage relative humidity range: 20 to 85%, non-condensing
Operating atmospheric pressure: 700 hPA to 1,060 hPA
Storage atmospheric pressure: 700 hPA to 1,060 hPA
Non-pyrogenic: Fluid pathway only
2 to 6 hours in 30-minute increments. Default is 4
hours.
0.05-30 U.
%, Units or Off. Default is Off.
30 minutes to 8 hours in 30-minute increments.
Maximum of 7. Cannot exceed the Maximum Bolus.
30 minutes to 2 hours.
10 to 50 units in 1-unit increments. Default is 10.0 U.
Off, or 1 to 24 hours in 1-hour increments. Default is
Off.
Rolling 90-day period.
English, French. Default is English.