PDM specifications
Size: 6.3 cm wide x 13.0 cm long x 1.0 cm high (2.48" x 5.1" x 0.39")
Weight: 106 grams (3.74 oz)
Screen active area: 10.2 cm diagonal ± 5% (4.0" ± 5%)
Operating temperature range: 5 °C to 40 °C (41 °F to 104 °F)
Storage temperature range: 0°C to 30 °C (32 °F to 86 °F)
Operating relative humidity range: 20% to 90%, non-condensing
Storage relative humidity range: 20% to 90%, non-condensing
Operating atmospheric pressure: 700 hPA to 1,060 hPA
Storage atmospheric pressure: 700 hPA to 1,060 hPA
Communication distance: The PDM and Pod should be
•
At start-up: Adjacent and touching, either in or out of the tray, to ensure
proper communication during priming.
•
During normal operation: Within 1.5 metres (5 feet) of each other.
Depending on the location, the communication distance may handle
separations of up to 15 metres (50 feet) away.
Waterproof rating: IP22 (avoid liquids)
Alarm type: Audible. Output: ≥ 45 db(A) at 1 metre
Notification type: Audible and vibratory
Battery: Rechargeable Li-ion battery, 3.7 V, 1,300 mAh
Only use the NUU mobile Li-Ion battery Model NUBA1 (Insulet PN 18363)
with the PDM.
Battery life: Full charge covers approximately 2 days of typical use after 2 years of
typical use
Battery charger operating line voltage: 100 to 240 VAC, 50/60 Hz
Only use the NUU mobile Switching Adapter Model HJ-0501000E1-US
(Insulet PN 18401) with the PDM.
Appendix
179
Effective Date: 18OCT2020, PCO-000235