POWER-PRO™ XT
Guidance and Manufacturer's declaration - Electromagnetic Immunity
The model 6500 Power-PRO™ XT cot is suitable for use in the electromagnetic environment specified below. The
customer or the user of the model 6500 Power-PRO™ XT cot should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrostatic fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
Note: U
is the alternating current mains voltage prior to application of the test level.
T
www.stryker.com

EMC Information

IEC 60601 Test level Compliance level
+6 kV contact
+6 kV contact
+8 kV air
+2 kV for power
+2 kV for power
supply lines
supply lines
+1 kV for input/
+1 kV for input/
output lines
output lines
+8 kV differential
+8 kV differential
mode
+2 kV common mode
+2 kV common
<5% U
T
(>95% dip in U
(>95% dip in U
)
T
for 0,5 cycle
for 0,5 cycle
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5 seconds
for 5 seconds
3 A/m
6500-109-005 REV A
Electromagnetic Environment Guidance
Floors should be wood, concrete, or
ceramic tile.
+8 kV air
synthetic material, the relative humidity
should be at least 30%.
Applies to:
• Cot,
• SMRT Charger (6500-201-010),
AC/DC Charger (6500-070-000),
• D
WALT
®
e
DC/DC Charger (6500-072-000).
• D
WALT
®
e
Main power quality should be that of a
typical commercial or hospital environment.
Applies to:
• SMRT Charger (6500-201-010),
AC/DC Charger (6500-070-000).
• D
WALT
®
e
Main power quality is that of a typical
mode
commercial and/or hospital environment.
Applies to:
• SMRT Charger (6500-201-010),
mode
AC/DC Charger (6500-070-000).
• D
WALT
®
e
<5% U
Main power quality should be that of a typical
T
)
commercial and/or hospital environment. If
T
the user of the charger requires continued
40% U
operation during power main interruptions,
T
)
it is recommended that the device be
T
powered from an uninterrupted power
70% U
supply or a battery.
T
)
T
<5% U
T
)
T
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial and/or hospital
environment.
Applies to:
• Cot,
• SMRT Charger (6500-201-010),
AC/DC Charger (6500-070-000),
• D
WALT
®
e
DC/DC Charger (6500-072-000).
• D
WALT
®
e
If floors are covered with
Return To Table of Contents
1-93
English