Guidance and Manufacturer's declaration - Electromagnetic Immunity

®
The GoBed
II MedSurg Bed, Model FL28EX is suitable for use in the electromagnetic environment specified below .
The customer or the user of the GoBed
environment .
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrostatic fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, voltage variations
and short interruptions on
power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a .c . mains voltage prior to applications of the test level .
T
www.stryker.com

EMC Information

®
II MedSurg Bed, Model FL28EX should assure that it is used in such an
IEC 60601 Test Level
+6 kV contact
+6 kV contact
+8 kV air
+8 kV air
+2 kV for power
+2 kV for power
supply lines
supply lines
+1 kV for input/
+1 kV for input/
output lines
output lines
+8 kV differential mode
+8 kV differential mode
+2 kV common mode
+2 kV common mode
<5%Ut (95% dip in Ut) for
<5%Ut (95% dip in Ut)
0 .5 cycle
for 0 .5 cycle
40%Ut (60% dip in Ut) for
40%Ut (60% dip in Ut)
5 cycles
for 5 cycles
70%Ut (30% dip in Ut) for
70%Ut (30% dip in Ut)
25 cycles .
for 25 cycles .
<5% Ut (>95% dip in Ut)
<5% Ut (>95% dip in Ut)
for 5 sec .
for 5 sec .
3 A/m
3 A/m
28EX-009-001 REV B
Compliance Level
Environment Guidance
Floors
concrete, or ceramic tile .
floors are covered with synthetic
material, the relative humidity
should be at least 30% .
Main power quality should be
that of a typical commercial or
hospital environment .
Main power quality is that of
a typical commercial and/or
hospital environment .
Main power quality should be
that of a typical commercial
and/or hospital environment .
If the user of the GoBed
MedSurg Bed, Model FL28EX
requires continued operation
during power main interruptions,
it is recommended that the
device be powered from an
uninterrupted power supply or
a battery .
Power
fields should be at levels
characteristic
location in a typical commercial
and/or hospital environment .
Electromagnetic
should
be
wood,
If
®
II
frequency
magnetic
of
a
typical
Return To Table of Contents
1-53