Guidance and Manufacturer's declaration - Electromagnetic Immunity
Isolibrium is suitable for use in the electromagnetic environment specified below. The customer or the user of
Isolibrium should make sure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrostatic fast transient/
burst
IEC 61000-4-4 *
Surge
IEC 61000-4-5 *
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11 *
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage before applications of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and
Isolibrium is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of Isolibrium can help prevent electromagnetic interferences by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and Isolibrium as recommended
below, according to the maximum output power of the communications equipment.
www.stryker.com

EMC Information

IEC 60601 Test Level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/ output
lines
+1 kV lines to lines
+2 kV lines to earth
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Isolibrium.
2971-209-001 REV B
Compliance Level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/ output
lines
+1 kV lines to lines
+2 kV lines to earth
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Main power quality should
be that of a typical
commercial or hospital
environment.
Main power quality should
be that of a typical
commercial or hospital
environment.
Main power quality should
be that of a typical
commercial and/or hospital
environment. If the user of
Isolibrium requires
continued operation during
power main interruptions, it
is recommended that the
device be powered from an
uninterrupted power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial and/or hospital
environment.
1-23