WARNING (CONTINUED)
Failure to properly clean or dispose of contaminated mattress or other cot components will increase the risk of
bloodborne pathogens and may cause injury to the patient or operator.
Escaping fluid under pressure can penetrate the skin causing serious injury. Avoid the hazard by relieving pressure
before disconnecting hydraulic or other lines. Tighten all connections before applying pressure. If an accident
occurs, see a doctor immediately. Any fluid injected into the skin must be surgically removed within a few hours or
gangrene may result. Doctors unfamiliar with this type of injury should reference a knowledgeable medical source.
To avoid the risk of injury, do not use bare hands to check for hydraulic leaks.
CAUTION
The cot can be set at any cot load height position. Establish the required cot load height before placing the cot
into service.
When charging a battery in an ambulance, locate the charger in an enclosed cabinet and out of patient reach
during transport.
Set the cot load height to the proper stop height prior to operation.
Installation of the safety hook should be done by a certified mechanic familiar with ambulance vehicle construction.
Consult the vehicle manufacturer before installing the safety hook and be sure that the installation of the safety
hook does not damage or interfere with the brake lines, oxygen lines, fuel lines, fuel tank or electrical wiring of
the vehicle.
Before operating the cot, clear any obstacles that may interfere and cause injury to the operator or patient.
Do not "jog" the cot past the established cot load height of the product when the safety bar engages the vehicle
safety hook or damage may occur to the product.
When unloading the cot from the patient compartment, ensure that the caster wheels are safely set on the ground
or damage to the product may occur.
Do not "jog" the cot past the load height while the safety bar is engaged.
Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours).
Only use the battery and charger as specified.
The cot is not for use with an AC adapter.
Ensure that the battery is fully charged prior to placing into service. An uncharged or depleted battery may cause
poor cot performance.
Ensure that the restraints are not entangled in the base frame when raising and lowering the cot.
Wheel lock(s) are only intended to help prevent the cot from rolling while unattended and to aid in patient transfer.
A wheel lock may not provide sufficient resistance on all surfaces or under loads.
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 40 pounds (18 kg).
To avoid damage to the equipment hook, the weight of the accessories or equipment must not exceed 35 pounds
(15.9 kg).
To avoid damage to the retractable head section oxygen bottle holder (if equipped), the weight of the equipment
must not exceed 40 pounds (18 kg).
Do not use two head end oxygen bottle holders at the same time.
Do not store items under the cot mattress. Storing items under the mattress can interfere with the operation of
the cot.
The weight of the equipment in the pocketed backrest storage pouch (if equipped) must not exceed 20 pounds
(9 kg).
The weight of the equipment in the head end storage flat (if equipped) must not exceed 40 pounds (18 kg).
DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE UNIT.
Maximum water temperature should not exceed 180°F/82°C.
Maximum air dry temperature (cart washers) should not exceed 240°F/115°C.
Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand held wand is being used to wash the
unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.
Towel dry all casters and interface points.
Failure to comply with these instructions may invalidate any/all warranties.
Always remove the battery before washing the cot.