TROUBlEShOOTING GUIDE (CONTINUED)
PROBlEM
Base does not retract in
the powered mode.
Base does not extend in the manual
mode.
Base does not retract in the manual
mode.
Litter does not retract in the manual
mode (with patient weight).
Litter does not extend in the manual
mode.
High Speed retract does not engage.
Return To Table of Contents
1-78
Troubleshooting Guide
1.
Check power indicator LED.
A. If blinking constant RED, change battery.
B. If blinking a patterned RED short, short, long flash:
i. Check for broken or disconnected wires.
ii. Check for 24 VDC at connector (B) on the main cable
by the motor while pressing the extend (+) button. If
voltage is present, replace (in order) the hall effect sensor,
solenoid, and or 'A' valve. If voltage is not present, go
to step iii.
iii. Check for 24 VDC on electronics assembly pins 2
white and 6 red on (F) while pressing the (+a) button. If
voltage is not present, replace the electronics assembly.
If voltage is present, replace the wire harness.
6500-109-005 REV A
SOlUTION
PAGE(S)
1-80,
1-81
www.stryker.com