POWER-PRO™ XT (CONTINUED)
The model 6500 Power-PRO™ XT cot is suited for use in the electromagnetic environment specified below. The
customer or the user of the model 6500
Immunity Test
Conducted
RF
150 kMz to 80 MHz
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
80 MHz to 2,5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the model 6500
used exceeds the applicable RF compliance level above, the model 6500
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the model 6500
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
www.stryker.com
EMC Information
P
ower-PRO™ XT cot should assure that it is used in such an environment.
IEC 60601
Compliance
Test level
level
3 Vrms
3 Vrms
3 V/m
3 V/m
P
ower-PRO™ XT cot.
6500-109-005 REV A
Electromagnetic Environment - Guidance
Portable
and
mobile
RF
equipment should be used no closer to any part
of the model 6500 Power-PRO™ XT cot, including
cables, than the recommended separation
distance calculated from the equation appropriate
for the frequency of the transmitter.
Recommended separation distance
d=1.2√P
Applies to:
• SMRT Charger (6500-201-010),
• D
WALT
®
AC/DC Charger (6500-070-000)
e
d=1.2√P
80 MHz to 800 MHz
d=2.3√P
800 MHz to 2.5 GHz
where
is the maximum output power rating
P
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
P