Guidance and manufacturer's declaration - electromagnetic emissions
The Model 4701 is intended for use in the electromagnetic environment specified below. The customer or the user of
the Model 4701 should assure that it is used in such an environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The Model 4701 is suitable for use in the electromagnetic environment specified below. The customer or the user of
the Model 4701 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
EN/IEC 61000-4-2
Electrostatic fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
www.stryker.com

EMC Information

Compliance
Group 1
Class A
Class A
Complies
EN/IEC 60601 test level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output lines
+8 kV differential mode
+2 kV common mode
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
) for
T
T
5 cycles
70% U
(30% dip in U
) for
T
T
25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec.
4701-209-001 REV B
Electromagnetic environment
The Model 4701 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The Model 4701 is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Compliance level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output lines
+8 kV differential mode
+2 kV common mode
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
) for
T
T
5 cycles
70% U
(30% dip in U
) for
T
T
25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec.
Electromagnetic
environment-guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Main power quality should
be that of a typical
commercial or hospital
environment.
Main power quality is that
of a typical commercial or
hospital environment.
Main power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Model 4701 requires
continued operation during
power main interruptions, it
is recommended that the
device be powered from an
1-47