POWER-PRO™ XT (CONTINUED)
The model 6500 Power-PRO™ XT cot is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the model 6500 Power-PRO™ XT cot can help prevent
electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the model 6500 Power-PRO™ XT cot as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Return To Table of Contents
1-94
EMC Information
Recommended separation distances between portable and mobile RF
communications equipment and the Power-PRO™ XT.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.17√P
0,12
0,38
1,2
3,8
12
m
80 MHz to 800 MHz
d=1.17√P
0,12
0,38
1,2
3,8
12
6500-109-005 REV A
800 MHz to 2.5 GHz
d=2.33√P
0,23
0,73
2,3
7,3
23
www.stryker.com