Emergency Stop - Agilent Technologies Turbo-V 750 TwisTorr Mode D'emploi

Table des Matières

Publicité

16
Instructions for Use
Use of Turbo-V 750/850 TwisTorr Pumping System

Emergency Stop

To immediately stop the Turbo-V 750/850 TwisTorr Pumping System
in an emergency condition it is necessary to remove the supply cable
from the mains plug. It must be however noted that this operation, in
addition to cutting off the pump power, also cuts off all other
controller functions, such as Purge and Controlled Vent management
and the capability of communicating with the system in which the
pump is integrated via I/Os, Serial or Profibus.
Furthermore, this operation may not ensure the immediate stopping
of the rotor. Rotation speed of the rotor will decrease according to
the degree of vacuum present in the system.
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
(A)
301/406

Publicité

Table des Matières
loading

Table des Matières