Agilent Technologies Turbo-V 750 TwisTorr Mode D'emploi page 317

Table des Matières

Publicité

Vibration level (displacement)
Lubricant
Cooling requirements
Coolant water
Noise level
Power supply Input for Turbo-V 750/850
TwisTorr Pumping System Power Supply
(969-6521):
Input voltage:
Input freq.:
Max input power:
Max operating power:
Power supply Input for Turbo-V 750/850
TwisTorr Pumping System:
Input voltage:
Input freq.:
Max input power:
Max output power:
International Protection of Pumping System
(Controller + Pump)
Protection fuse for power supply system
969-6521
Compliance with:
Power cable
Serial communication (T-Plus kit)
Installation category
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
< 0.01 μm at inlet flange
Permanent lubrication
Forced air (up to 35 °C ambient temperature)
Water (mandatory if ambient temperature > 35 °C)
Minimum flow: 100 l/h (0.89 GPM)
Temperature: +15 °C to +30 °C
Pressure: 3 to 5 bar (45 to 75 psi)
FAN off = < 52 dB(A) at 1 meter
FAN on = < 55 dB(A) at 1 meter
100 - 240 Vac (± 10 %)
50 - 60 Hz
550 VA
450 W
48 Vdc (± 10 %)
DC
450 W
400 W (Ramp status)
300 W (Normal/Autotuning status) with water cooling
200 W (Normal/Autotuning status) with forced air cooling
IP54
2 x T8A250H
UNI EN 292-1
UNI EN 292-2
EN 61010-1 (2001)
EN 61326-1 (2006)
Emission level : Class B for Residential Application
Immunity level : for Industrial Application
With European or NEMA plug 3 meters long (optional)
RS232 cable with a 9-pin D type male connector and a 9-pin
D type female connector, and T-Plus software (optional)
II
(A)
Technical Information
Technical Specification
17
317/406

Publicité

Table des Matières
loading

Table des Matières