WARNING!
If the pump is fastened on a vacuum chamber with different flange Agilentts
there exists the chance of twisting or tearing-off in case the rotor is suddenly
blocked. The combination of different flange types is not allowed in any case.
Agilent Vacuum will not accept any liability for all damages resulting from
not allowed fastening.

Installation of ISO-K Flange

For ISO-K flange connections, fix the two flanges with the clamps as
shown in the following figure (a protective screen can optionally be
used).
Figure 75
Use the required number of claw clamps:
Tighten the claw clamps with a torque of 25 Nm.
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
10 for ISO 160 flange
18 for ISO 200 flange
Technical Information
Connection of the High Vacuum Flange
(A)
17
383/406