17
Technical Information
Output Power during High Temperature Condition
Output Power during High Temperature Condition
During operation the controller limits the power provided to the
motor according to:
WARNING!
The user must set the gas type load (WIN 157) and Cooling mode (WIN 106)
before starting the pump.
For details on the applied power derate see the following figures.
The applied power limit is readable by serial line (WIN 155).
If the body temperature exceed 55 °C or the Controller temperature
exceed 65 °C, the controller goes in status 6 (Fail).
336/406
Gas type load (selectable WIN 157: see the following "Window
Meanings" paragraph)
Cooling mode (selectable WIN 106: see the following "Window
Meanings" paragraph)
Body temperature (can be read by WIN 222).
Electronics temperature (can be read by WIN 216).
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
(A)