Instructions for Use
How to Stop the System
To stop the pump you can use one of the following methods:
1
2
3
Emergency Stop
To immediately stop the Turbo-V 3K-G in an emergency condition it
is necessary to remove the mains. It must be however noted that this
operation, in addition to cutting off the pump power, also cuts off all
other controller functions, such as Purge and Controlled Vent
management and the capability of communicating with the system in
which the pump is integrated via I/Os, Serial or Profibus.
Furthermore, this operation may not ensure the immediate stopping
of the rotor. Rotation speed of the rotor will decrease according to
the degree of vacuum present in the system.
296/360
unplugging the provided mating I/O connector
giving a remote stop signal through the I/O connector (see para.
"J1-REMOTE I/O" of annex "Technical Information")
giving a remote stop signal via the serial RS 232/485 interface
(see para. "RS232/RS 485 Communication Description" of annex
"Technical Information").