Instructions for Use
Use of Turbo-V 750/850 TwisTorr Pumping System
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal. Do not
use the pump in presence of explosive gases. The pump is designed to pump
high throughput of N
heavier than Ar, please contact Agilent technical support for information.
How to Start the Turbo-V 750/850 TwisTorr
Pumping System
Before starting the system, please check that the mating I/O
connector is unplugged. If the system is connected to a remote I/O,
make sure the stop signal (see para. "J1-REMOTE I/O" of annex
"Technical Information") is given.
To start the system please follow the following steps:
1
2
3
4
The integrated controller automatically recognizes the interlock and
start signals presence and start up the pump.
298/406
, Ar and lighter gas. Should you need to pump gases
2
unplug (if present) the system mating I/O connector;
to switch on the Turbo-V 750/850 TwisTorr Pumping System it is
necessary to supply a voltage of 48 Vdc using an external Power
Supply (969-6521);
pump the vacuum chamber down to 0.1 mbar
give the Turbo-V 750/850 TwisTorr Pumping System a start signal
by one of the following methods:
a
connecting the provided mating I/O connector
b
giving a remote start signal through the I/O connector (see
para. "J1-REMOTE I/O" of annex "Technical Information")
c
giving a remote start signal via the serial RS 232/485
interface (see para. "RS232/RS 485 Communication
Description" of annex "Technical Information").
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
(A)