NOTE: When using CAPIOX FX25 Oxygenator module and Hardshell
Reservoir separately,
a) Hold the module firmly and separate it from Hardshell Reservoir by
removing connecting ring.
b) Set reservoir in the holder (Code No.:XX*CXH18R, 801804 or 801139)
for CAPIOX FX25 Oxygenator with Hardshell Reservoir and pull the lever
down. (see Fig. 1)
c) Set the oxygenator module in its holder (Code No.:XX*CXH25F) and close
the "lock" cover. (see Fig. 3- )
b. CAPIOX FX25 Oxygenator module without Hardshell Reservoir:
Set the holder at the appropriate height to prevent the arterial line and the blood
cardioplegia line from kinking.
Set oxygenator module in its holder (Code No.:XX*CXH15) (See Fig.3- )
CAUTION If the product is dropped during set-up, do not use it. Replace with
another device.
3. Set the sampling system into the holder.
4. Connect water lines (1/2" (12.7 mm) tubing or couplers) to the water ports of
CAPIOX FX25. (See Fig. 4-(1))
CAUTION Use the upper port for water supply and the lower port for water
drainage. Otherwise heat exchange will not be performed
sufficiently.
5. Start water circulation through heat exchanger and circulate for at least 5 minutes.
Check for leaks.
WARNING Do not use an oxygenator that leaks.
6. Attach 1/2" (12.7 mm) venous line to the venous blood inlet port of venous reservoir.
(See Fig. 4-(2))
7. Attach 3/8" (9.5 mm) arterial line to the blood outlet port of CAPIOX FX25
Oxygenator. (See Fig. 4-(3))
8. Remove the cap from the blood cardioplegia port and attach 1/4" (6.4 mm) blood
cardioplegia line tubing. (See Fig. 4-(4)).
WARNING If the port is not to be used, push cap in and twist it to seal tightly or
attach a short 1/4" (6.4 mm) tubing and clamp for security.
9. Connect 3/8" (9.5 mm) pump line to the blood outlet port of Hardshell Reservoir.
Connect other end to 3/8" (9.5mm) blood inlet port of oxygenator. (See Fig. 4-(5))
10. The blood cardioplegia line must be connected to the luer port lateral to the blood
outlet port of the oxygenator if oxygenated blood is needed for blood cardioplegia
(see Fig. 4-(4)).
11. Band all connections in the circuit.
12. Connect 1/4" (6.4 mm) gas line to gas inlet port. (See Fig. 4-(6))
WARNING DO NOT OBSTRUCT GAS OUTLET PORT. (See Fig. 4-(7))
13. Thermistor probes can be used with Y.S.I. 400 series* temperature monitor by use
of the following parts.
Blue Cable (venous line): Code No.: CX*BP021
Red Cable (arterial line): Code No.: CX*BP022
* Y.S.I. 400 (trade name: Measurement specialties Inc.)
14. Connect suction lines and vent line to suction ports on Hardshell Reservoir. Suction
ports of CAPIOX FX25 Hardshell Reservoir have blue caps. Remove the blue caps
before connection. (See Fig. 5)
15. For use of CAPIOX FX25 Oxygenator module without Hardshell Reservoir, connect
venous side male luer lock of sampling system to venous line. (See Fig. 6)
16. When the luer port located on venous blood inlet port of CAPIOX FX25 Hardshell
Reservoir is to be used, attach a three-way stopcock before use.
17. Check figures and labels to make sure tubing is connected to the correct port.
Special care should be taken to connect the correct tubing to the 1/4" (6.4 mm)
gas inlet port (Fig. 4-(6)), and 1/4" (6.4 mm) blood cardioplegia port (Fig. 4-(4)).
CAUTION Ensure that all connected parts including the luer caps, the lock
adapters and the port caps are securely affixed. Loose connections
may cause contamination or a blood leak.
8
Fig. 3
Holder (XX*CXH25F)
Lock cover
(When using the
oxygenator module
of the CX*FX25RW
separately)
Fig. 4
Venous blood inlet port
(2)
Venous line
Reservoir
outlet port
(7)
Gas outlet
port
(5)
Blood outlet port
(5)
(to pump)
Line from pump
Blood inlet port
Blood cardioplegia port
Fig. 5
Fig. 6
Venous side stopcock
(blue spot)
Arterial side stopcock
(red spot)
Holder (XX*CXH15)
(When using
CX*FX25E)
Blood cardioplegia line
Water inlet
line
(1)
Water ports
(1)
(4)
Water outlet
(6)
line
(3)
Gas line
Gas inlet port
Blood outlet port
Arterial line
Suction line
Vent line
Suction port
Venous side male
luer lock