OPERATING THE UNIT

Suction control

Recommendations on suction settings for lithotripsy using BSC
handpieces and probes are provided separately.
Suction flow can be adjusted by using the roller clamp on the
suction tubing.
Excessive suction can impair the endoscopic view, collapse the
organ, or damage the mucosa. Always check that the suction tubing
is inserted into the pinch valve and that the pinch valve is functioning
properly.
Make sure that suction is always operational. Continuous suction
during ultrasound lithotripsy is required to evacuate fragments and
liquids as well as to ensure the cooling of the probe and the handpiece.
Interruption of the suction must be avoided. It is operator's
responsibility to continuously monitor the suction flow to check that it
is operational and that the level of lower pressure is correctly set.
Fragments blocked in the lumen of the probe and the handpiece
may lead to loss of suction and heating of the probe. If blockage occurs,
stop lithotripsy to remove fragments from the probe and from the
handpiece lumen before continuing.