Technical Data - Boston Scientific Swiss LithoClast Select Mode D'emploi

Masquer les pouces Voir aussi pour Swiss LithoClast Select:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TECHNICAL DATA

DESCRIPTION
Distributor
Manufacturer
Model
Power supply (voltage)
Power supply (frequency)
Power consumption
Ultrasound frequency
Ultrasound power consumption
Energy output
Classification
Operating mode
Primary fuse
Compressed air supply
Weight
Housing dimensions
Operating conditions
44
FB-411_US_ed_2013_10_Not_Swiss LithoClast Select BSC.indd 44
Boston Scientific Corporation
Marlborough, MA USA 01752
EMS SA, CH – 1260 Nyon, Switzerland
Swiss LithoClast
Select System
®
100 – 240 VAC
50 – 60 Hz
500 VA
23 – 26.4 kHz
78 W
Appro. 85 mJ
EN-60601-1: Class I
Applied part BF
IP 20 (Foot pedal: IP X8)
MDD 93/42 EEC: Class IIb
Continuous modes
Ø 5 X 20 mm T4A 250 V, time lag, low
breaking capacity
3.5 to 6.5 bar (3500 to 6500 hPa)
(51 psi to 94 psi)
11 kg (24.3 lb)
Height: 135 mm (5.3 in)
Width: 360 mm (14.2 in)
Depth: 420 mm (16.5 in)
Temp. +10°C to +30°C (50°F to 86°F)
Rel. humidity 30% to 75%
Atm. pressure 700 hPa to1060 hPa
(10.1 psi to 15.3 psi),
Altitude max. 3000 m (9800 ft)
Storage and transport conditions
-10°C to +40°C (14°F to 104°F)
10% to 95% relative humidity
500 hPa to 1060 hPa air pressure
(7.2 psi to 15.3 psi)
6/3/2015 2:33:58 PM