Ultrasound Power; Duty Cycle; Activation - Boston Scientific Swiss LithoClast Select Mode D'emploi

Masquer les pouces Voir aussi pour Swiss LithoClast Select:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Ultrasound power

The left rotary knob is used to adjust the ultrasound power from 10 %
to 100 % (in 10 % increments).

Duty cycle

The duty cycle is the ratio between ultrasound ON and ultrasound
OFF over an interval of time of 80 milliseconds. Set the duty cycle
using the arrow keys.
The last setting of the duty cycle is kept in memory at power off.
The handpiece and probe error detection is possible only with
a duty cycle set at 100 % (see Troubleshooting chapter).

Activation

Press the ultrasound (left) foot pedal halfway to activate the
suction (pinch valve opens).
Press the ultrasound (left) foot pedal completely to activate the
ultrasound handpiece and probe with suction.
10
FB-411_US_ed_2013_10_Not_Swiss LithoClast Select BSC.indd 10
6/3/2015 2:33:52 PM

Publicité

Table des Matières
loading

Table des Matières