Disposable Stone Catcher device
The disposable Stone Catcher device is provided sterile. Read the
instruction manual delivered with it before performing installation.
Connect the short silicone tubing end with the conical connector to the
external suction source.
The pinch valve default state is closed and opens when the
ultrasound (left) foot pedal is depressed halfway.
Install a roller clamp on the suction tubing to enable control of
the suction flow independently of the lower pressure.
The sterile disposable Stone Catcher device cannot be
reprocessed.
4
FB-411_US_ed_2013_10_Not_Swiss LithoClast Select BSC.indd 4
Stone Catcher device tubing
The use of other tubing than the one supplied with the Stone
Catcher device can either create an incomplete occlusion of the tubing
in the pinch valve, or the tubing can collapse during use. Use only the
Stone Catcher device tubing provided by BSC.
The suction tube (standard suction tubing with 3 m length and 5 mm
internal diameter) must be sterile and single use. Use of a transparent
tubing allows for observation of fragment transport to visualize a
possible obstruction.