MAINTENANCE
CLEANING THE INTELECT® MOBILE
2
With the system disconnected from the power
source, clean the system with a clean, lint-free
cloth moistened with water and mild antibacterial
soap. If a more sterile cleaning is needed, use a
cloth moistened with an antimicrobial cleaner.
Cleaning should be performed daily.
Do not submerge the system in liquids. Should the
unit accidentally become submerged, contact the
dealer or DJO Service Department immediately.
Cleaning the LCD Screen
Clean the LCD with a clean, dry cloth, in the same
way as cleaning the computer monitor screen. Do
not use abrasive materials or chemicals or liquids.
Cleaning instruction for the Ultrasound applicator
The sound head may be cleaned with alcohol
between each therapy session. The Aluminium
surface may be disinfected with alcohol, but avoid
the plastic area.
CALIBRATION REQUIREMENTS
The unit was calibrated during the manufacturing
process and doesn't need calibration during the
product life.
INSTRUCTION FOR SOFTWARE
UPGRADE
1.
Go to the Chattanooga website
www.chattanoogarehab.com
2.
Go to Intelect® Mobile 2 product page
3.
Complete the registration form to be
informed about new product software
version availability and IFU updates (if not
already done before)
4.
Go to the downloads tab
5.
Download firmware upgrade zip file and
extract the file
6.
Erase the USB drive supplied with the
Intelect® Mobile 2
INTELECT
MOBILE 2 USER MANUAL
®
7.
Copy the extracted files on to the USB
drive
8.
Switch OFF the device
9.
Insert USB key drive into the USB port on
the back of the device
10.
Switch ON the device
11.
Device will automatically detect firmware
update availability and commence
upgrade, the upgrade will take some
minutes and the power must not be
switched off during the upgrade
12.
Once firmware update is finished, Home
screen will be displayed and the USB drive
can be removed. Device is ready for use.
13.
Check software version in settings
DEVICE DISPOSAL
Council Directive 2012/19/EU concerning Waste
Electrical and Electronic Equipment (WEEE)
requires not to dispose of WEEE as municipal
waste. Contact your local distributor for
information regarding disposal of the unit and
accessories.
In case of incident occurring in relation to the
device, contact local Health Authority and the
manufacturer:
DJO France S.A.S. for Chattanooga product range
Centre Européen de Fret
3 rue de Bethar
64990 Mouguerre
France
internationalproductsupport@djoglobal.com
E N