GENERAL WARNINGS AND PRECAUTIONS
• This unit should be operated at +5°C to +40°C and 15% to 90% Relative Humidity. The unit should
be transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity.
• Use of parts or materials other than DJO's can degrade minimum safety.
• Connect to this unit only items and equipment that have been specified in this IFU as part of the
ME SYSTEM or that have been specified as being compatible with the ME SYSTEM.
• DO NOT disassemble, modify, or remodel the unit or accessories. This may cause unit damage,
malfunction, electrical shock, fire, or personal injury.
• DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, including, but not
limited to, inflammables, water, and metallic objects, to prevent unit damage, malfunction, electrical
shock, fire, or personal injury.
• Before each use, inspect Ultrasound Applicator for cracks, which may allow the ingress of
conductive fluid.
• Before each use, inspect Applicator cables, STIM cables and associated connectors.
• Before each use, inspect Vacuum Electrode Cups and Lead Hoses for cracks and damage which
may not allow the vacuum to properly secure the electrodes.
• Handle Ultrasound Applicator with care. Inappropriate handling may adversely affect its
characteristics.
• Caution should always be exercised with current densities more than 2mA/cm².
• There are no user-serviceable parts inside the unit. If a malfunction occurs, discontinue use
immediately and consult dealer for repair service.
• In case of device unused with battery embedded, it is recommended to connect the device at least
once every 4 months to allow battery recharge.
• For waveforms with a DC component:
» Do not shave electrodes application area
» Warn the patient that tingling sensation under electrodes is normal and it is not linked
to burn risk.
» Rinse thoroughly treatment area with tap water immediately after the treatment
INDICATIONS FOR USE
CAUTION
INTELECT
MOBILE 2 USER MANUAL
®
11