Télécharger Imprimer la page

Chattanooga Intelect Mobile 2 ULTRASOUND Manuel D'utilisation page 27

Masquer les pouces Voir aussi pour Intelect Mobile 2 ULTRASOUND:

Publicité

E N
WAVEFORMS
Advice on size and type of electrodes to be used is given in device User Interface treatment guidelines.
CC: Constant Current
CV: Constant Voltage
IFC (Interferential) Traditional (4 Pole)
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Interferential Current is a medium frequency
waveform. Current is distributed through two
channels (four electrodes). The currents cross each
other in the body at the area requiring treatment.
The two currents interfere with each other at
this crossing point, resulting in a modulation of
the intensity (the current intensity increases and
decreases at a regular frequency).
Output Mode
Electrodes
Available on Channel
1&2,
Treatment Time
1-60 Minutes
Mode Selection
CC
Output Intensity
0-100 mA (CC)
Beat Frequency
1-200 Hz
Carrier Frequency
2000-10,000 Hz
Cycle Time
Continuous or User Defined
Sweep Time
14 sec
Sweep Low Beat Frequency
1-199 Hz
Sweep High Beat Frequency
2-200 Hz
Scan Percentage
Static, 40%, 100%, Manual
IRMS
0-78mA
DC component
No
SYSTEM
TENS- Asymmetrical Biphasic
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
The Asymmetrical Biphasic waveform has a short
pulse duration. It is capable of strong stimulation
of the nerve fibers in the skin as well as of muscle
tissue. This waveform is often used in TENS devices.
Because of its short pulse, the patient typically
tolerates the current well, even at relatively high
intensities.
Output Mode
Electrodes
Output Intensity
0-140 mA (CC) 0-140 V
(CV)
Available on Channel
1,2
Treatment Time (Stim)
1-60 minutes
Treatment Time (Combo)
1-30 minutes
Mode Selection (Stim)
CC or CV
Mode Selection (Combo)
CV
Amplitude Modulation
0% (off) to 100% on 10%
steps
Burst Frequency
0-10 bps
Cycle Time
Continuous or User Defined
Frequency
1-200 pps
FrequencySweep
On/Off
Phase Duration
30-400 µsec
Sweep time
14 sec
Sweep Low Frequency
1-199 pps
Sweep High Frequency
2-200 pps
IRMS
0-50mA
DC component
No
INTELECT
MOBILE 2 USER MANUAL
®
27