ULTRASOUND SPECIFICATIONS
Frequency
Duty Cycles
Pulse Repetition Rate
Pulse duration:
OUTPUT POWER
1cm2
1MHz
3MHz
1
0.9
2W
1.8W
3W
2.7W (*)
2W/cm2
2W/cm2
3W/cm2
3W/cm2
(*) An error of + 0.25 W can be measured with 1cm2 US applicator, pulse mode 100Hz at 10% or 20% Duty Cycle.
Unless otherwise specified, ultrasound
controls accuracy is:
± 20 %.
Peak to Average Ratio:
1:1, at 50% Duty Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
Beam Nonuniformity Ratio<5:1
Beam Type
Collimating
Treatment Time
1 to 30 min
1 MHz; 3 MHz
10%, 20%, 50%, Continuous
16, 48, or 100 Hz
1 -31.25 ms
Max (ON): 31.25 ms
Min (OFF): 5ms
2cm2
5cm2
1MHz
3MHz
1MHz
1.5
1
2.5
3W
2W
5W
4.5W
3W
7.5W
2W/cm2
2W/cm2
2W/cm2
3W/cm2
3W/cm2
3W/cm2
SYSTEM
10cm2
3MHz
1MHz
2.7
6
5.4W
12W
8.1W
18W
2W/cm2
2W/cm2
3W/cm2
3W/cm2
INTELECT
MOBILE 2 USER MANUAL
®
25
3MHz
6.8
6.8W
13.6W
1W/cm2
2W/cm2