38
SYSTEM
WAVEFORM GUIDANCE
Recommended parameters for each waveform in each of the indications
Waveform
Default settings
TENS- Asymmetrical
80Hz; cont; 20min
Biphasic
TENS- Symmetrical Biphasic 50Hz; cont; 20min
TENS - HAN
HAN settings (fixed); 30min
50Hz;2s ramp;
VMS™
5s/5s;20min
50Hz;2s ramp;
VMS™ Burst
5s/5s;20min
50Hz; 500ms burst; 1:3;
VMS™ FR
20min
IFC (Interferential)
4000Hz; 80/150Hz;
Traditional (4 Pole)
auto 40%; cont; 20min
IFC Premodulated
4000Hz; 80/150Hz;
(Traditional 2 Pole)
cont; 20min
Isoplanar Vector
4000Hz; 80/150Hz;
Interferential Current
cont; 20min
Dipole Vector Interferential
4000Hz; 80/150Hz;
Current
45°/6s; cont; 20min
2500Hz; 50Hz; DC 50%;
Russian
cont; 20 min
High Voltage Pulsed Current
100Hz; cont; 20min
(HVPC)
MONOPHASIC: Monophasic
1.0ms/500ms; 5min
Rectangular pulsed
MONOPHASIC: Monophasic
1.0ms/500ms; 5min
Triangular pulsed
SURGED:
50Hz; 1ms; 15 min
Monophasic Rectangular
SURGED:
50Hz; 1ms; 15 min
Monophasic Triangular
Diadyn settings (fixed),
Diadynamic Waveforms
10 min
Träbert settings (fixed),
Träbert (Ultrareiz)
10min
Microcurrent
Neg; 1Hz; 20min
GALVANIC: Continuous
Pos; 10min
Low Level Galvanic
Pos; 1mA; 40min
GALVANIC: Interrupted
Pos; 10 min
INTELECT
MOBILE 2 USER MANUAL
®
Management and alleviation of post operative pain
Freq
Intensity
strong sensation/
80-150Hz
no pain
strong sensation/
80-150Hz
no pain
-
-
strong sensation/
80-150Hz
no pain
strong sensation/
80-150Hz
no pain
-
-
use default settings and increase intensity to strong sensory
level/no pain
use default settings and increase intensity to strong sensory
level/no pain
use default settings and increase intensity to strong sensory
level/no pain
use default settings and increase intensity to strong sensory
level/no pain
-
-
use default settings and increase intensity to strong sensory level/no pain
-
-
-
-
-
-
-
-
Use default Diadynamic settings and increase intensity to strong sensory level/
no pain
Use default Träbert settings and increase intensity strong to sensory level/no pain
Use default settings and increase intensity to 40-100µA
-
-
-
-
-
-
E N
Duty cycle
Treatment Time
continuous
when in pain
continuous
when in pain
-
-
continuous
when in pain
continuous
when in pain
-
-
when in pain
when in pain
when in pain
when in pain
-
-
-
-
-
-
-
-
-
-
when in pain/
multiple hours
-
-
-
-
-
-

Publicité