ELECTROTHERAPY PATIENT PREPARATION AND
ELECTRODE PLACEMENT
• Examine the skin for any wounds and
clean the skin.
• Apply the electrodes to the treatment
area.
• Ensure the electrodes are applied
securely to the skin.
• Ensure good contact between each
electrode and the skin.
• Check the electrode contact regularly
during the treatment.
• Examine the skin again after the
treatment.
• Choose electrodes that fit the
anatomy.
• View the Electrode Placement
recommendations in the Treatment
Review screen for the particular
modality being used for treatment
as a reference point only prior to
administering treatment.
• Follow electrode manufacturer
instructions.
• Please note the smaller the electrode
size the higher the current density.
DURA-STICK® Electrodes
DURA-STICK® Electrodes are a self adhesive,
disposable product designed specifically for use
with Intelect® Mobile 2.
It is recommended that DURA-STICK® Electrodes
be used whenever possible to ensure the highest
level of contact with the treatment area and most
uniform delivery of the prescribed electrotherapy
treatment.
For Electrotherapy operation refer to page 42
DURA-STICK® Electrode Instructions
PATIENT PREPARATION
Connecting Lead Wires
Insert the lead with the Red (+) electrode
connector into one DURA-STICK® Electrode.
Insert the lead with the Black (-) electrode
connector into the other electrode.
Make certain the lead wires are seated completely
into the electrodes.
NOTE: Use of conductive medium or sponges is not
required or recommended. DURA-STICK
manufactured to ensure the optimum conductivity during
therapy when properly applied.
Securing Electrodes
Remove the DURA-STICK® Electrodes from the
protective backing.
Apply to the treatment area as prescribed.
Ensure the entire electrode surface is in contact
with patient skin by pressing into place.