Télécharger Imprimer la page

Chattanooga Intelect Mobile 2 ULTRASOUND Manuel D'utilisation page 30

Masquer les pouces Voir aussi pour Intelect Mobile 2 ULTRASOUND:

Publicité

30
SYSTEM
WAVEFORMS (CONTINUED)
IFC Premodulated (Traditional 2 Pole)
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Premodulated Current is a medium frequency
waveform. Current comes out of one channel (two
electrodes). The current intensity is modulated: it
increases and decreases at a regular frequency (the
Amplitude Modulation Frequency).
Output Mode
Electrodes
Available on Channel
1, 2
Treatment Time (STIM)
1-60 Min
Treatment Time (COMBO)
1-30 Min
Mode Selection
CC or CV
Output Intensity
0-100 mA (CC) 0-100 V
(CV)) Carrier
Beat Fixed (Sweep Off)
1-200 Hz
Cycle Time
Continuous or User Defined
Frequency
2,000-10,000 Hz
Sweep Low Beat Frequency
1- 199 Hz
Sweep High Beat Frequency
2-200 Hz
IRMS
0-55mA
DC component
No
Russian
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Russian Current is a sinusoidal waveform, delivered
in bursts or
series of pulses.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC or CV
Output Intensity
0-100 mA (CC) 0-100 V
(CV)
Burst Frequency
1-100 bps
Carrier Frequency
2,500 Hz
Cycle Time
Continuous or User Defined
Duty Cycle
10%, 20%, 30%, 40%,
50%
Ramp
0-5 sec
IRMS
0-39mA
DC component
INTELECT
MOBILE 2 USER MANUAL
®
VMS™ Burst
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
VMS Burst is a symmetrical biphasic waveform
delivered in a burst format. Because the pulse is
relatively short, the waveform has a
low skin load, making it suitable for applications
requiring high intensities, such as muscle re-
education protocols.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC or CV
Output Intensity
0-140 mA (CC) 0-140 V
(CV)
Anti-
Off or On
Burst Frequency
1-200 bps
Channel Mode
Single, Reciprocal, Co-
Contract Phase
Cycle Time
Continuous or User Defined
Duration
30-400 µsec
Ramp
0-5 sec
Set Intensity
Individual/both Channel
Intensity
Setting in Reciprocal and Co-Contract modes
IRMS
0-50mA
DC component
No
MONOPHASIC: Monophasic Rectangular
Pulsed
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
The Monophasic Rectangular Pulsed waveform is an
interrupted unidirectional current with a rectangular
pulse shape.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Phase Duration
0.1-1,000 ms
Phase Interval
5-5,000 ms
IRMS
0-47mA
DC component
Yes
E N

Publicité

loading