1 0
INDICATIONS FOR USE
ULTRASOUND INDICATIONS
INDICATIONS
• Musculoskeletal conditions facilitating
pain relief to the affected area.
CONTRAINDICATIONS
• Do not use for symptomatic local pain
relief unless etiology is established
or unless a pain syndrome has been
diagnosed.
• Do not use when cancerous lesions
are present in the treatment area.
• Do not use when patient is suspected
or known to have infectious disease
and/or disease where it is advisable,
for general medical purposes, to
suppress heat or fevers.
• Do not use over or near bone growth
centers until bone growth is complete.
• Do not use over the thoracic area
if the patient is using a cardiac
pacemaker.
• Do not use over a healing fracture.
• Do not use over or applied to the eye.
• Do not use over a pregnant uterus.
• Tissue necrosis might result if the
device is used on ischemic tissues
in individuals with vascular disease,
where the blood supply would not
keep up with the metabolic demand.
• Do not use Intelect® Mobile 2 on
patients who have or have had
implantable neurostimulating cardiac
demand pacemakers, ICD, or other
implantable electronic devices.
INTELECT
MOBILE 2 USER MANUAL
®
• Do not use Intelect® Mobile 2 on
patients with body worn electro
mechanical medical devices, i.e. insulin
pump.
• Do not use this system in an MRI or
CT environment. The Intelect® Mobile
2, its components, and accessories
are not to be present in an MRI or CT
environment.
ADDITIONAL PRECAUTIONS
Additional precautions should be used when
ultrasound is used on patients with the following
conditions:
• Over an area of the spinal cord
following a laminectomy, i.e., when
major covering tissues have been
removed
• Over anesthetic areas
• On patients with hemorrhagic
diatheses
ADVERSE EFFECTS
• Stinging sensation and temporary
local sensitivity have been reported
during or following ultrasound
therapy.
E N