GENERAL WARNINGS AND PRECAUTIONS
• This device should be used only under the continued supervision of a physician or licensed
practitioner.
• Contaminated sponges, electrodes, leadwires, and gel can lead to infection.
• Use of electrode on multiple patients can lead to infection.
• Do not apply electro stimulation treatment during bath, shower, sauna,..
• DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of an ultrasonic
diathermy system.
• DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of any microware and RF
shortwave diathermy system.
• DO NOT operate this unit in an environment where other devices are being used that intentionally
radiate electromagnetic energy in an unshielded manner.
• Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate
properly when electrical stimulation is in use.
• Simultaneous connection of a PATIENT to a high frequency surgical ME EQUIPMENT may result in
burns at the site of the STIMULATOR electrodes and possible damage to the STIMULATOR.
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to
any part of the Intelect® Mobile 2, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
• Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
• Battery replacement by inadequately trained personnel could result in fire or explosion. Please
read carefully the battery replacement instructions in the Mobile 2 IFU before attempting to
replace the battery.
• Device is designed to comply with electromagnetic safety standards. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with
instructions for use, may cause harmful interference to other devices in the vicinity. Harmful
interference to other devices can be determined by turning this equipment on and off. Try to
correct the interference using one or more of the following:
» Reorient or relocate the receiving device
» Increase the separation between the equipment
» Connect the equipment to an outlet on a different circuit from that to which the other
device(s) are connected
» Consult your authorized DJO dealer for help.
• Disconnect the system from the power source before attempting any maintenance, installation,
removal, or replacement procedures to prevent electrical shock and possible damage to system.
• The Intelect® Mobile 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±6
kV when first grasping the Ultrasound applicator. In the event of such a discharge, the Intelect®
Mobile 2 may display a permanent error. The Intelect® Mobile 2 will terminate all active outputs
(stim, ultrasound,), automatically place the unit in a safe state.
• To prevent Electro-Static Discharge (ESD) at greater than ±6 kV:
» Grasp and hold the Ultrasound prior to starting treatment. If the applicator must be put