SYSTEM
GENERAL SYSTEM OPERATING AND STORAGE
TEMPERATURE
Operating Conditions
The device will meet its requirement under the following conditions:
Temperature:
5°C to 40°C
Relative Humidity:
Atmospheric Pressure:
Transport and Storage Conditions
The device will remain in proper condition under the following conditions:
Temperature:
-20°C to 60°C
Relative Humidity:
Atmospheric Pressure:
Time required for the Intelect® Mobile 2 to warm from the minimum storage
temperature between uses until the Intelect® Mobile 2 is ready for its INTENDED USE
when the ambient temperature is 20 °C: 5h
Time required for the Intelect® Mobile 2 to cool from the maximum storage
temperature between uses until the Intelect® Mobile 2 is ready for its INTENDED USE
when the ambient temperature is 20 °C: 5h
IPXX Rating for Unit
Rated to IP21
IP2* Protection against fingers or other object not greater than 80mm in length and
12mm in diameter
*1 Protection from vertically dripping water
IPXX Rating for US applicator
Rated to IPX7
IPX7 Protection from immersed in water (up to 1m depth)
RED
RF transmitter/receiver characteristics:
- Frequency Band transmission:
- Modulation type:
- Data rate:
- Effective radiated power: +6dBm
INTELECT
MOBILE 2 USER MANUAL
®
15% to 90%
70kPa to 106kPa
10% to 90%
50kPa to 106kPa
2400–2483.5 MHz
GFSK
up to 2Mbps 500kHz deviation at 2Mbps