E N
WAVEFORMS (CONTINUED)
MONOPHASIC: Monophasic Triangular
Pulsed
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
The Monophasic Triangular Pulsed waveform is an
interrupted unidirectional current with a triangular
pulse shape.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-60 mA (CC)
Phase Duration
0.1-1,000 ms
Phase Interval
5-5,000 ms
IRMS
0-27mA
DC component
Yes
GALVANIC: Continuous
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Galvanic Current is a direct current flowing in one
direction only. The current can be continuous or
interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-40 mA (CC)
Cycle Time
Continuous, or User Defined
Polarity Reversal
On or Off
With Polarity Reversal On, Polarity will change in
the middle of the treatment time.
IRMS
0-44mA
DC component
Yes
SYSTEM
GALVANIC: Interrupted
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Galvanic Current is a direct current flowing in one
direction only. The current can be continuous or
interrupted.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-40 mA (CC)
Pulse Duration
136 µsec
Phase Interval
25 usec
Polarity Reversal
On or Off
With Polarity Reversal On, Polarity will
change in the middle of the treatment time.
Polarity Reversal Ramp
1 sec
IRMS
0-41mA
DC component
Yes
Träbert (Ultrareiz)
The Träbert Current is a monophasic waveform
with a phase duration of 2 ms and a pause of 5 ms
resulting in a frequency of approximately 143Hz.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-80 mA (CC)
Frequency
143 pps
Polarity Reversal
On or Off
With Polarity Reversal On, Polarity will change
in the middle of the treatment time.
Phase Duration
2 ms
IRMS
0-47mA
DC component
Yes
INTELECT
MOBILE 2 USER MANUAL
®
31